Electromagnetic Compatibility Product Specifications
9 - 26 046-004667-00 A7™ Operating Instructions
9.19 Electromagnetic Compatibility
The A7 meets the requirements of IEC 60601-1-2 / EN 60601-1-2.
NOTE: Using accessories, sensors and cables other than those specified may
result in increased electromagnetic emission or decreased
electromagnetic immunity of the equipment.
NOTE: The anesthesia machine or its components should not be used adjacent
to or stacked with other equipment. If adjacent or stacked use is
necessary, the anesthesia machine or its components should be
observed to verify normal operation in the configuration in which it will
be used.
NOTE: The anesthesia machine needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC
information provided below.
NOTE: Other devices may interfere with this equipment even though they
meet the requirements of CISPR.
NOTE: When the input signal is below the minimum amplitude provided in
technical specifications, erroneous measurements could result.
NOTE: Use of portable or mobile communications devices can degrade the
performance of the equipment.
GUIDANCE AND DECLARATION - ELECTROMAGNETIC EMISSION
The A7 is intended for use in the specified electromagnetic environment. The customer or the user of the
A7 should assure that it is used in such an environment as described below.
EMISSIONS TEST COMPLIANCE ELECTROMAGNETIC ENVIRONMENT - GUIDANCE
Radio frequency (RF)
emissions CISPR 11
Group 1 The A7 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Radio frequency (RF)
emissions CISPR 11
Class B A7 is suitable for use in all establishments, including
domestic establishments and those directly connected to
the public low-voltage power supply network that supplies
buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
Not applicable
TABLE 9-39
Guidance and Declaration - Electromagnetic Emission

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
BrandMindray
ModelA7
CategoryMedical Equipment
LanguageEnglish

Related product manuals