A7™ Operating Instructions 046-004667-00 7 - 17
Maintenance Cleaning and Disinfection
7.9.7 PAW Gauge
1. The PAW gauge is a component that is plugged into position for use. It is not necessary to
remove this component to clean it. However, if removal is desired, simply hold it and lift it
straight up from the breathing system block (see FIGURE 7-14).
FIGURE 7-14 PAW Gauge Removal
CAUTION: Do not autoclave the following components: Paw gauge, flow sensor,
and bellows. These components cannot withstand immersion or the
heat and pressure of autoclaving.
2. Clean the PAW gauge with a soft, lint-free cloth and a recommended cleaning agent (see
TABLE 7-2 on page 7-8 and TABLE 7-3 on page 7-9). Allow it to dry thoroughly.
3. Re-insert the PAW gauge if it was removed. Prior to use after cleaning or disinfecting, power
up the system and follow the on-screen prompts to perform the leak test and the
compliance test (see section 4.5 (page 4-8) "Leak and Compliance Tests").

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
TypeAnesthesia Workstation
BrandMindray
ModelA7
Display15-inch color touchscreen
Oxygen FlushYes
Gas InletsO2, N2O, Air
Ventilation ModesSIMV, PSV
Breathing SystemIntegrated
Gas MonitoringO2, CO2, N2O
Monitoring ParametersFiO2
CO2 MonitoringMainstream or Sidestream
Power Supply100-240 V, 50/60 Hz

Related product manuals