Start Mechanical Ventilation Operations
5 - 18 046-004667-00 A7™ Operating Instructions
5.6 Start Mechanical Ventilation
NOTE: Before starting a new mechanical ventilation mode, ensure that all
related ventilation parameters are set appropriately.
To start mechanical ventilation from the Standby mode:
1. Set the Auto/Manual ventilation switch to Manual.
2. Exit Standby by touching the main screen or by turning on the fresh gas.
3. Set the Auto/Manual ventilation switch to Auto. The A7 System begins mechanical
ventilation.
5.7 Stop Mechanical Ventilation
To stop mechanical ventilation:
1. Ensure that the breathing system is set up and the APL valve is set properly before stopping
mechanical ventilation.
2. Set the Auto/Manual ventilation switch to Manual. The A7 System stops mechanical
ventilation.
5.8 Relationships of Ventilation Parameters
Ventilation modes may share the same ventilation parameters and values. For example, SIMV-VC
and VCV both include Vt, Plimit, Rate, Tpause, and PEEP. Therefore, these parameter values that
are linked may be passed from the previous ventilation mode to the current mode. Section C.9
‘Linked Ventilation Parameter’ on page C-11 includes a table that lists how the linked parameter
values are set when changing ventilation modes.
Unlinked ventilation parameter values are set according to relationship equations. Section C.10
Ventilation Parameter Relationships’’ on page C-13 includes a table of equations to show how
unlinked parameter values are set when changing ventilation modes.
5.9 Parameter Monitoring (Numerics)
The system displays parameter monitored values in the monitored parameter area. The
monitored parameters are separated into three groups: pressure, volume and gas (available with
the AG module).
5.9.1 Pressure
The Pressure parameter group consists of three(3) parameters:
Airway Peak Pressure (PEAK)
Plateau Pressure (PLAT) or Mean Pressure (MEAN)
Positive End Expiratory Pressure (PEEP)
If the parameter data is out of range, it displays as “– – –”.
NOTE: The high alarm limit for Airway Peak Pressure (PEAK) displays to the top
right of the reading. The low alarm limit for Airway Peak Pressure
(PEAK) displays to the bottom right of the reading.
NOTE: Configure the display of either Plateau Pressure (PLAT) or Mean
Pressure (MEAN) from the System menu tab.

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
BrandMindray
ModelA7
CategoryMedical Equipment
LanguageEnglish

Related product manuals