Alarm and Prompt Messages Alarms and Messages
6 - 14 046-004667-00 A7™ Operating Instructions
If the calculated alarm limit is more than the high threshold of setting range or less than the low
threshold, the corresponding threshold is used as the auto alarm limit.
6.5.4 Setting CO
2
Apnea Delay Time (software bundle version
02.09.00 and later)
The Apnea CO
2
alarm is triggered when no breath is detected within a specified time.
To set the CO
2
Apnea Delay Time:
1. On the main screen, select the Alarms softkey.
The Alarms window displays.
2. Select the Limits tab. (see FIGURE 6-6)
3. Select the CO2 Apnea Delay Time button and set it to 10 sec, 15 sec, 20 sec, 25 sec, 30 sec,
35 sec, or 40 sec.
6.6 Alarm and Prompt Messages
This section lists the following alarms and messages:
• Physiological Alarm Messages
• Technical Alarm Messages
• Prompt Messages
For each alarm message, corresponding actions are given instructing you to troubleshoot problems.
If the problem persists, contact your service personnel.
NOTE: The Disable in Manual and Cardiac Bypass mode column indicates how
this alarm is controlled by the alarm on/off button and the cardiac
bypass mode button in manual mode.
NOTE: The Disabled in Standby mode column indicates which physiological
alarms will be automatically disabled in the Standby mode.
NOTE: The Disabled in Monitor mode column indicates which physiological
alarms will be automatically disabled in the Monitor mode.