A7™ Operating Instructions 046-004667-00 2 - 7
Installation Install the Vaporizer
2.3.1 Filling and Draining the Vaporizer
Install the vaporizers with a Selectatec interlock system that are compliant to ISO 80601-2-13 on
the A7 unit. See section A.9 (page A-4) "Vaporizers". Refer to the vaporizer manufacturers
Instructions For Use for filling or draining the vaporizer and other information.
WARNING: Ensure that the correct anesthetic agent is used. The vaporizer is
designed with the specific anesthetic agent named on it and further
indicated by color coded labelling. The concentration of the anesthetic
agent actually output will vary if the vaporizer is filled with the wrong
agent.
WARNING: Do not reuse the agent drained from the vaporizer. Treat as a hazardous
chemical and follow local regulations for proper disposal.

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
TypeAnesthesia Workstation
BrandMindray
ModelA7
Display15-inch color touchscreen
Oxygen FlushYes
Gas InletsO2, N2O, Air
Ventilation ModesSIMV, PSV
Breathing SystemIntegrated
Gas MonitoringO2, CO2, N2O
Monitoring ParametersFiO2
CO2 MonitoringMainstream or Sidestream
Power Supply100-240 V, 50/60 Hz

Related product manuals