General System Overview Product Description
1 - 2 046-004667-00 A7™ Operating Instructions
1.1 General System Overview
1.1.1 General Description
The A7 Anesthesia System is a device intended to administer, continuously or intermittently, a
general inhalation anesthetic to a patient, and to maintain a patient's ventilation. The A7 also
provides for ventilatory monitoring of the patient. The anesthesia system is intended to be used in
the patient environment.
The A7 Anesthesia System consists of a main unit (includes an anesthetic ventilator, electronic flow
control system, and backup flow control system) and a detachable breathing system. The applied
parts of the anesthesia system are breathing tubes and masks. Connect the patient to the anesthesia
system via breathing tubes and masks.
The A7 Anesthesia System provides the following ventilation modes:
Volume Control Ventilation (VCV), which includes the Pressure Limit Ventilation (PLV) function
Pressure Control Ventilation (PCV) with/without Volume Guarantee (VG) ventilation mode
Synchronized Intermittent Mandatory Ventilation (SIMV) with VC mode
(with/without PS option)
Synchronized Intermittent Mandatory Ventilation (SIMV) with PC mode
(with/without PS option)
Pressure Support (PS) ventilation mode
Spontaneous ventilation in Manual mode with APL fully open.
Manual Ventilation through the use of a breathing bag.
Cardiac Bypass mode
•Monitor Mode
•ACGO Mode
Electronic PEEP is available in all automatic ventilation modes. User control of the inspiratory flow rise
(Tslope) is possible in PCV, SIMV, and PS modes. Automatic fresh gas compensation limits the effect
on the patient ventilation from changes in fresh gas flow rate by the operator. The traditional bellows
system is driven by oxygen and makes patient disconnections clearly visible.
The A7 Anesthesia System provides the following common functions:
Automatic leak detection
Circuit gas leakage compensation and automatic compliance compensation
Cylinder and central pipeline gas supply connections available for gas input
Electronically adjustable PEEP
Electronic timer to display the duration between the start and end of an operation
Top light available
External connect on to a patient monitor
•Network upgrade
Flow trigger mode available for PS and SIMV
Auxiliary O
2
and air supply
Active AGSS or optional Passive scavenging
•N
2
O cutoff
Cardiac Bypass alarm mode

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
BrandMindray
ModelA7
CategoryMedical Equipment
LanguageEnglish

Related product manuals