A7™ Operating Instructions 046-004667-00 8 - 7
AG and O2 Concentration Monitoring Make AG Settings
8.7 Make AG Settings
Perform the settings below when the anesthesia system is configured with an external AG
module.
8.7.1 Set CO
2
Unit
To change the CO
2
Unit:
1. Select Setup softkey > System tab.
2. Select the CO
2
Unit button.
3. Choose between mmHg, kPa and %.
4. Select the Accept button to confirm the change, or select the Cancel button to disregard
the change.
8.7.2 Set CO
2
Placement
To change the CO
2
Placement:
1. Select Setup softkey > Display tab.
2. Select the CO
2
Placement button.
3. Choose between To p and Bottom.
4. Select the Accept button to confirm the change, or select the Cancel button to disregard
the change.
8.7.3 Set CO
2
Scale
To change the CO2 Scale:
1. Select Setup softkey > Display tab.
2. Select the CO
2
Scale button.
3. Choose the desired scale.
4. Select the Accept button to confirm the change, or select the Cancel button to disregard
the change.
8.7.4 Gas Bench Flow Rate
To change the Gas Bench Flow Rate:
1. Select Setup softkey > General tab.
2. Select the Gas Bench Flow Rate button.
3. Choose between High (recommended), Med and Low, as follows:
High: 200 ml/min for high volume watertrap; 120 ml/min for low volume watertrap
Med: 150 ml/min for high volume watertrap; 90 ml/min for low volume watertrap
Low: 120 ml/min for high volume watertrap; 70 ml/min for low volume watertrap
4. Select the Accept button to confirm the change, or select the Cancel button to disregard
the change.