Ventilation Modes Parameters and Factory Defaults
C - 10 046-004667-00 A7™ Operating Instructions
Trigger —Range:
1 to 15 L/min
Step: 1
Defaults:
Adult: 3 L/min
Pediatric: 2 L/min
Infant: 2 L/min
—Range:
1 to 15 L/min
Step: 1
Defaults:
Adult: 3 L/min
Pediatric: 2 L/min
Infant: 2 L/min
Range:
1 to 15 L/min
Step: 1
Defaults:
Adult: 3 L/min
Pediatric: 2 L/min
Infant: 2 L/min
Tslo pe Range: 0.0 to 2.0 sec
Step: 0.1 sec
Default: 0.2 sec
Range: 0.0 to 2.0 sec
Step: 0.1 sec
Default: 0.2 sec
Range: 0.0 to 2.0 sec
Step: 0.1 sec
Default: 0.2 sec
Range: 0.0 to 2.0 sec
Step: 0.1 sec
Default: 0.2 sec
PS —Range: On, Off
Step: —
Default: Off
—Range: On, Off
Step: —
Default: Off
——
Bypass —— Range: On, Off
Step: —
Default: Off
Alarm —— Range: On, Off
Step: —
Default: On
Apnea Ti Range: 0.2 to 5.0 sec
Step: 0.1 sec
Default:
5.0 sec (adult)
3.0 sec (Pediatric)
2.0 sec (Infant)
Monitor —— Range: On, Off
Step: —
Default: Off
ACGO —— Range: On, Off
Step: —
Default: Off
PARAMETER VCV SIMV-VC PCV SIMV-PC PS MANUAL

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
TypeAnesthesia Workstation
BrandMindray
ModelA7
Display15-inch color touchscreen
Oxygen FlushYes
Gas InletsO2, N2O, Air
Ventilation ModesSIMV, PSV
Breathing SystemIntegrated
Gas MonitoringO2, CO2, N2O
Monitoring ParametersFiO2
CO2 MonitoringMainstream or Sidestream
Power Supply100-240 V, 50/60 Hz