A7™ Operating Instructions 046-004667-00 C
-9
Parameters and Factory Defaults Ventilation Modes
Pinsp —— Range:
PEEP+5 to 70 cmH
2
O
Step: 1 cmH
2
O
Defaults:
Adult: 15 cmH
2
O
Pediatric: 10 cmH
2
O
Infant: 10 cmH
2
O
Range:
PEEP+5 to 70 cmH
2
O
Step: 1 cmH
2
O
Defaults:
Adult: 15 cmH
2
O
Pediatric: 10 cmH
2
O
Infant: 10 cmH
2
O
——
Tpause Range:
Off, 5% to 60%
Step: 1%
Default: 10%
Range:
Off, 5% to 60%
Step: 1%
Default: 10%
——
Plimit Range:
10 to 100 cmH
2
O
Step: 1 cmH
2
O
Defaults:
Adult: 50 cmH
2
O
Pediatric: 40 cmH
2
O
Infant: 20 cmH
2
O
Range:
10 to 100 cmH
2
O
Step: 1 cmH
2
O
Defaults:
Adult: 50 cmH
2
O
Pediatric: 40 cmH
2
O
Infant: 20 cmH
2
O
——
PlimVG Range: 5 - 100
cmH
2
O
Step: 1 cmH
2
O
Default:
Pinsp
——
PEEP Range:
Off, 3 to 30 cmH
2
O
Step: 1 cmH
2
O
Default: Off
Range:
Off, 3 to 30 cmH
2
O
Step: 1 cmH
2
O
Default: Off
Range:
Off, 3 to 30 cmH
2
O
Step: 1 cmH
2
O
Default: Off
Range:
Off, 3 to 30 cmH
2
O
Step: 1 cmH
2
O
Default: Off
Range:
Off, 3 to 30 cmH
2
O
Step: 1 cmH
2
O
Default: Off
ΔP —Range:
3 to 50 cmH
2
O
Step: 1
Defaults:
Adult: 8 cmH
2
O
Pediatric: 5 cmH
2
O
Infant: 5 cmH
2
O
—Range:
3 to 50 cmH
2
O
Step: 1
Defaults:
Adult: 8 cmH
2
O
Pediatric: 5 cmH
2
O
Infant: 5 cmH
2
O
Range:
3 to 50 cmH
2
O
Step: 1
Defaults:
Adult: 8 cmH
2
O
Pediatric: 5 cmH
2
O
Infant: 5 cmH
2
O
PARAMETER VCV SIMV-VC PCV SIMV-PC PS MANUAL