Ventilation Modes Parameters and Factory Defaults
C - 8 046-004667-00 A7™ Operating Instructions
PARAMETER VCV SIMV-VC PCV SIMV-PC PS MANUAL
Vt Range:
20 to 1500 ml
Step: 1
Defaults:
Adult: 600 ml
Pediatric: 120 ml
Infant: 20 ml
Range:
20 to 1500 ml
Step: 1
Defaults:
Adult: 600 ml
Pediatric: 120 ml
Infant: 20 ml
——
VtG —— Range:
Off, 20 to 1500 ml
Step: 1
Default:
Off
——
Rate Range: 4 to 100 bpm
Step: 1 bpm
Defaults:
Adult: 8 bpm
Pediatric: 15 bpm
Infant: 20 bpm
Range: 4 to 100 bpm
Step: 1 bpm
Defaults:
Adult: 8 bpm
Pediatric: 15 bpm
Infant 20 bpm
Range: 4 to 100 bpm
Step: 1 bpm
Defaults:
Adult: 8 bpm
Pediatric: 15 bpm
Infant: 20 bpm
Range: 4 to 100 bpm
Step: 1 bpm
Defaults:
Adult: 8 bpm
Pediatric: 15 bpm
Infant: 20 bpm
——
Min. Rate Range: 2 to 60 bpm
Step: 1 bpm
Defaults:
Adult: 4 bpm
Pediatric: 6 bpm
Infant: 12 bpm
I:E Range: 4:1 to 1:8
Step: 0.5
Default: 1:2
Range: 4:1 to 1:8
Step: 0.5
Default: 1:2
——
Tinsp Range: 0.2 to 5 sec
Step: 0.1 sec
Defaults:
Adult: 2.0 sec
Pediatric: 1.0 sec
Infant: 1.0 sec
Range: 0.2 to 5 sec
Step: 0.1 sec
Defaults:
Adult: 2.0 sec
Pediatric: 1.0 sec
Infant: 1.0 sec
——