A7™ Operating Instructions 046-004667-00 4 - 15
Preoperative Tests Leak and Compliance Tests
Time out
Example:
Test result cannot be shown due to an internal
communication error.
Select Retry to repeat the Automatic Circuit Leak Test &
Compliance test. In certain rare instances, after pressing
"Retry" the manual circuit leak test will be performed, but
the progress bar will be blank and will not show that the
manual leak test was completed. If this anomaly occurs,
press the Cancel button on the screen to continue.
or
Select Accept to proceed.
Backup Flow Control Enabled
Example
For the A7 System, the backup flow control is active.
Follow on-screen instructions to troubleshoot the
problem.
or
Select Accept to proceed to operational mode using
previous compliance values.
RESULTS COMMENTS/OPTIONS
TABLE 4-4 Manual Circuit Leak Test Results

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
TypeAnesthesia Workstation
BrandMindray
ModelA7
Display15-inch color touchscreen
Oxygen FlushYes
Gas InletsO2, N2O, Air
Ventilation ModesSIMV, PSV
Breathing SystemIntegrated
Gas MonitoringO2, CO2, N2O
Monitoring ParametersFiO2
CO2 MonitoringMainstream or Sidestream
Power Supply100-240 V, 50/60 Hz

Related product manuals