A7™ Operating Instructions 046-004667-00 xv
Warnings Introduction
WARNING: Make sure that all connections are reliable. Any leak in the system can
result in erroneous readings due to patient breathing gas mixed with
ambient air.
WARNING: When using the AG module to perform AG measurements on the
patients who are receiving or have recently received anesthetic agents,
connect the outlet to the sample gas return port to prevent the medical
staff from breathing in the anesthetic agents.
WARNING: The operation of the A7 below the minimum amplitude or value
provided in technical specifications may cause inaccurate results.
WARNING: Using accessories, sensors and cables other than those specified may
result in increased electromagnetic emission or decreased
electromagnetic immunity of the equipment.
WARNING: The A7 or its components should not be used adjacent to or stacked
with other equipment. If adjacent or stacked use is necessary, the A7 or
its components should be observed to verify normal operation in the
configuration in which it will be used.
WARNING: Other devices may interfere with this equipment even though they
meet the requirements of CISPR.
WARNING: Use only accessories specified in this manual. Using other accessories
may cause incorrect measured values or equipment damage.
WARNING: Disposable accessories cannot be reused. Reuse may degrade
performance or cause cross-contamination.
WARNING: Check the accessories and their packages for damage. Do not use them
if any sign of damage is detected.
WARNING: At the end of its service life, the equipment, as well as its accessories,
must be disposed of in compliance with the guidelines regulating the
disposal of such products, and in accordance with local regulations for
contaminated and biologically hazardous items.
WARNING: Ensure that the correct anesthetic agent is used. The vaporizer is
designed with the specific anesthetic agent named on it and further
indicated by color coded labelling. The concentration of the anesthetic
agent actually output will vary if the vaporizer is filled with the wrong
agent.
WARNING: Do not reuse the agent drained from the vaporizer. Treat as a hazardous
chemical and follow local regulations for proper disposal.
WARNING: The suction shall be used with a suction catheter.
WARNING: Do not modify this equipment without authorization of the
manufacturer.
WARNING: Connecting electrical equipment to MSO(multiple socket-outlets)
effectively leads to creat an ME(medical electrical) system, and can
result in a reduced level of safety.
WARNING: The anesthesia system shall not be serviced or maintained while being
connected on a patient.
WARNING: Additional MULTIPLE SOCKET- OUTLET or extension cord shall not be
connected to the ME SYSTEM.

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
BrandMindray
ModelA7
CategoryMedical Equipment
LanguageEnglish

Related product manuals