Introduction Notes
xx 046-004667-00 A7™ Operating Instructions
NOTE: The A7 battery supply is not a user serviceable component. Only an
authorized service representative can replace the battery supply. If the
system is not used for an extended period, contact a service
representative to have the battery supply disconnected. The batteries
may be subject to local regulations regarding disposal. At the end of
the battery life, dispose of the battery supply in accordance with local
regulations.
NOTE: Areas designated for the servicing of oxygen equipment shall be clean,
free of oil and grease, and not used for the repair of other equipment.
NOTE: Opening the cylinder valve quickly may cause unexpected pressure
differentials and create a potential for fire or explosion arising from
oxygen pressure shocks. Open and shut the cylinder valve slowly.
NOTE: Accuracy of the flowrate may be affected by varying inlet pressure,
varying outlet resistance, or varying ambient temperature.
NOTE: The power device, terminal units and pipeline system can be supplied
by one or several different manufacturers.
NOTE: Regional or national regulations that apply to manufacturers of
medical devices can exist.
NOTE: The measured values displayed on the screen are measured under BTPS
conditions.
NOTE: For the method of connecting A7 to external monitor or other devices,
please see Anesthesia Machine Bracket Installation Instructions.
NOTE: No repair should ever be attempted by anyone not having experience
in the repair of devices of this nature. Replace damaged parts with
components manufactured or sold by Mindray. Then test the unit to
ensure that it complies with the manufacturer’s published
specifications.
NOTE: The A7 can be equipped with one scavenger system to provide the best
match with the hospital’s waste-gas disposal system. The scavenger
system shall comply with ISO 80601-2-13.
NOTE: The A7 can be campatible with O2, N2O, air, halothane, enflurane,
isoflurane, sevoflurane and desflurane.
NOTE: Any patches will be deployed by trained engineering personnel at the
customer site.
NOTE: The Anesthesia System is compatible with gases (O2, N2O, and Air) and
anaesthetic agents (Halothane, Enflurane, Isoflurane, Sevoflurane, and
Desflurane).
NOTE: The leakage of AGSS is measured by the method recommended in ISO
80601-2-13.

Table of Contents

Other manuals for Mindray A7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A7 and is the answer not in the manual?

Mindray A7 Specifications

General IconGeneral
BrandMindray
ModelA7
CategoryMedical Equipment
LanguageEnglish

Related product manuals