5.2 System Check
Three system check modes are available: power-on self-test, leak test, and vaporizer test (for A9).
NOTE
Both the leak test and vaporizer test are used to check for leakage in the backend
circuit of the flowmeter. The difference lies in whether the anesthesia machine is
equipped with a vaporizer. During the system check, the anesthesia machine
automatically performs detection and selection. You can also manually select the
corresponding test on the standby screen. If you select the vaporizer test but no
electronic vaporizer is loaded, the test may fail.

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA