8-6 
8.8 Disassemble the Inspiration and Expiration Connectors 
8.8.1 Prepare for Disassembly 
8.8.1.1 Tools 
During parts disassembly and replacement, the following tools may be required: 
  None. 
8.8.1.2 Preparations 
Before disassembly, 
  Make sure that the anesthesia machine is turned off and disconnected from the A/C power 
source. 
  Bleed down the gas pressure inside the anesthesia machine to avoid personal injury or 
equipment damage. 
  Disconnect all pipeline and cylinder gas supplies. 
  Maneuver the anesthesia machine to an appropriate location and then apply the brake. 
8.8.2 Remove the Inspiration and Expiration Connectors 
Turn the locking nuts of the inspiration and expiration connectors counterclockwise. Pull out the 
inspiration and expiration connectors together with their locking nuts. Then remove the 
inspiration and expiration connectors. 
     
8.9 Disassemble the Water Collection Cup 
8.9.1 Prepare for Disassembly 
8.9.1.1 Tools   
During parts disassembly and replacement, the following tools may be required: 
  None. 
8.9.1.2 Preparations 
Before disassembly, 
  Make sure that the anesthesia machine is turned off and disconnected from the A/C power 
source. 
  Bleed down the gas pressure inside the anesthesia machine to avoid personal injury or 
equipment damage. 
  Disconnect all pipeline and cylinder gas supplies. 
  Maneuver the anesthesia machine to an appropriate location and then apply the brake. 
8.9.2 Remove the Water Collection Cup 
Turn the water collection cup counterclockwise to remove it.