Mindray A9 Service Manual

Mindray A9
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #268 background imageLoading...
Page #268 background image
7-44
Message
Priority
Cause
Solution
Range
temperature of the
MiniMedi CO2 module
is beyond the normal
range.
2. If the problem persists, replace the
CO2 module.
CO2 Check Cal. Low
The MiniMedi CO2
module fails.
1. Re-plug the CO2 module.
2. Recalibrate the CO2 module.
3. If the problem persists, replace the
CO2 module.
7.3.14 NMT Module Runtime Alarms
Message
Priority
Cause
Solution
NMT No Main
Cables
Low
The NMT main cable is
disconnected.
Check the connection between the
main cable and the NMT module.
NMT No Sensor Low
The NMT sensor is
disconnected.
1. Check the connection between the
sensor and the NMT main cable to
ensure normal connection.
2. If the alarm persists, replace the
sensor.
NMT Stimulation
Electrode Off
Low
The NMT stimulation
electrode falls off.
1. Check the connection between the
stimulation cable and the main cable to
ensure normal connection.
2. If the alarm persists, check the
connection between the stimulation
electrode and the patient.
NMT Sensor Comm.
Error
Low
An NMT sensor
communication error
occurs.
1. Check the connection between the
sensor and the NMT main cable to
ensure normal connection.
2. If the alarm persists, replace the
NMT module or sensor.
3. If the problem persists, check the
connection cable between the
backplane and the display.
4. If the problem persists, replace the
module rack (refer to 11.1.14).
5. If the problem persists, replace the
cable between the backplane and
display.
NMT Stimulation
Current Over Limit
Low
The output stimulation
current of the NMT
module exceeds the
limit.
1. Check the connection between the
stimulation cable and the main cable to
ensure normal connection between the
stimulation electrode and the patient.
2. If the alarm persists, replace the
NMT module or related component.
NMT Comm.
Abnormal
High
NMT communication is
abnormal.
1. Re-plug the NMT module, main
cable, sensor, and stimulation
electrode, or plug the NMT module to
another machine. Eliminate the error
situations one by one.
2. If the alarm persists, replace the
NMT module or related component.
3. If the problem persists, check the

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA