7-27
7.3.6 Electronic Vaporizer Runtime Alarms
Liquid Level Low
liquid level is lower
than 50 ml and higher
than or equal to 20
ml.
1. Observe the remaining anesthetic
agent through the glass tube. If the
anesthetic agent is insufficient, fill
the anesthetic agent.
2. If the anesthetic agent is
sufficient, reinstall the electronic
vaporizer.
3. If the problem persists, replace
the electronic vaporizer.
4. After the electronic vaporizer is
replaced and inserted, conduct the
power-on selftest, leak test, and
electronic vaporizer test (refer to
Almost Empty
liquid level is lower
than 20 ml.
1. Observe the remaining anesthetic
agent through the glass tube. If the
anesthetic agent is insufficient, fill
anesthetic agent.
2. If the anesthetic agent is
sufficient, reinstall the electronic
vaporizer.
3. If the problem persists, replace
the electronic vaporizer.
4. After the electronic vaporizer is
replaced and inserted, conduct the
power-on selftest, leak test, and
electronic vaporizer test (refer to
is Open
Information
2. Non-standby
and non-active
state:
Information
(first 2 min),
Low (2 min
later)
3. Non-standby
and active state:
Information
(first 2 min),
Med (2 min
open.
1. If the vaporizer lid is open, close
it.
2. Reinstall the electronic vaporizer.
3. If the problem persists, replace
the electronic vaporizer.
4. After the electronic vaporizer is
replaced and inserted, conduct the
power-on selftest, leak test, and
electronic vaporizer test (refer to
5.7).
Failure
boil-off space
temperature too high
(2) AA vaporizer
boil-off space
temperature sensor
Rectify the fault based on the alarm
error code by referring to 7.3.6.2.