11-3
11.1.2 Cylinder Yoke Assembly
No. Name P/N Test Item After Replacement Remarks
1
O2 cylinder yoke assembly
(electronic monitoring)
FRU
115-066820-00
5.4.2 O2 cylinder supply test.
Observe whether the O2
cylinder pressure is displayed
on the small display.
/
N2O cylinder yoke
assembly (electronic
monitoring) FRU
115-066821-00
5.4.1 N2O cylinder supply
test. Observe whether the N2O
cylinder pressure is displayed
on the small display.
/
Air cylinder yoke assembly
(electronic monitoring)
FRU
115-066822-00
5.4.3 Air cylinder supply test.
Observe whether the air
cylinder pressure is displayed
on the small display.
/

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA

Related product manuals