EasyManuals Logo

Mindray A9 Service Manual

Mindray A9
424 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #337 background imageLoading...
Page #337 background image
8-37
8.41 Disassemble the Air Filter
8.41.1 Prepare for Disassembly
8.41.1.1 Tools
During parts disassembly and replacement, the following tools may be required:
Phillips screwdriver
8.41.1.2 Preparations
Before disassembly,
Make sure that the anesthesia machine is turned off and disconnected from the A/C power
source.
Maneuver the anesthesia machine to an appropriate location and then apply the brake.
8.41.1.3 Pre-disassembly
1. Refer to 8.1.4Remove the Work Surface Cover Plate FRU to remove the work surface
cover plate.
2. Refer to step 1 in 8.23.2Remove the Patient Circuit Assembly to remove the patient circuit.
3. Refer to 8.28Disassemble the Volume Exchanger Assembly to remove the volume
exchanger assembly.
4. Refer to 8.29Disassemble the VE Box Assembly to remove the VE box assembly.
8.41.2 Remove the Air Filter
Pull out the air filter.
8.42 Disassemble the AG Module
8.42.1 Prepare for Disassembly
8.42.1.1 Tools
During parts disassembly and replacement, the following tools may be required:
Phillips screwdriver
8.42.1.2 Preparations
Before disassembly,
Make sure that the anesthesia machine is turned off and disconnected from the A/C power
source.
Maneuver the anesthesia machine to an appropriate location and then apply the brake.
Air Filter

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA