11-12
11.1.9 Drawer and Battery Box Assemblies
No. Name P/N Test Item After
Replacement
Remarks
1
Battery box
assembly (two
batteries)
115-066846-00
Power on the machine, access
the diagnostic maintenance
screen, select Data
Monitors > Power System,
and check the battery quantity
and voltage.
Includes two
batteries.
2
Upper-middle
drawer assembly
service part
115-066841-00 /
The drawer
assembly and
the left and right
rails on the cart
are included.
3
Lower drawer
assembly service
part
115-066842-00 /
The drawer
assembly and
the left and right
rails on the cart
are included.
4
Rail of the drawer
031-000664-00
/
/
4

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA

Related product manuals