7-41 
BIS SQI R<15%  Low 
prompting the user that 
the BIS value is not 
1. Check the contact between the BIS 
sensor and the patient skin. 
2. If the problem persists, replace the 
BIS SQI R<50%  Low  SQI R is low. 
1. Check the contact between the BIS 
sensor and the patient skin. 
2. If the problem persists, replace the 
BIS Sensor Expired  Low  The BIS sensor expires. 
1. Check the validity period of the BIS 
sensor. 
2. If the problem persists, replace the 
BIS Electrode 
Unconnected 
Low 
The BIS sensor 
electrode is not 
connected. 
1. Check whether the BIS sensor 
electrodes are properly connected. 
2. If the problem persists, replace the 
BISx Disconnected  Low 
The BIS cable is not 
connected to the plug-in 
module, or 
a communication error 
occurs. 
1. Check the connector between the 
BIS cable and the plug-in module. 
2. If the problem persists, replace the 
BIS cable. 
3. If the problem persists, replace the 
BIS Wrong Sensor 
Type 
Low 
A Mindray undeclared 
BIS sensor is used. 
1. Check the type of the BIS sensor. 
2. If the problem persists, replace the 
BIS Sensor Failure  Low 
positive electrode error 
2. If the problem persists, replace the 
Disconnect/Reconnect 
BIS 
Low 
and two positive 
electrode errors occur in 
total. The BIS module 
needs to be powered off 
and then powered on. 
The BIS module or 
BIS plug-in box needs to 
1. Re-plug the BIS sensor. 
2. If the problem persists, replace the 
BIS sensor. 
BIS Self Test Error  Low 
An error occurs during 
BIS module self-test. 
1. Re-plug the BIS module. 
2. If the problem persists, replace the 
 
7.3.13 CO2 Module Runtime Alarms 
CO2 Comm Stop  High 
A CO2 module error or 
communication error 
occurs. 
1. Re-plug the CO2 module, or replace 
the slot. 
2. If the problem persists, replace the 
CO2 module. 
3. If the problem persists, check the 
connection cable between the 
backplane and the display. 
4. If the problem persists, replace the