7-42
module rack (refer to 11.1.14).
5. If the problem persists, replace the
cable between the backplane and
CO2 Sensor High
Temp
Low
temperature is greater
1. Re-plug the CO2 module.
2. If the problem persists, replace the
CO2 Sensor Low
Temp
Low
temperature is less than
1. Re-plug the CO2 module.
2. If the problem persists, replace the
CO2 High Airway
Press.
Low
The airway pressure is
greater than 790 mmHg.
1. Check whether the airway pressure
is above 790 mmHg.
2. Reconnect the sampling line.
3. If the problem persists, replace the
CO2 Low Airway
Press.
Low
The airway pressure is
less than 428 mmHg.
1. Check whether the airway pressure
is below 428 mmHg.
2. Reconnect the sampling line.
3. If the problem persists, replace the
CO2 High
Barometric
Low
is greater than 790
1. Re-plug the CO2 module.
2. If the problem persists, replace the
CO2 Low
Barometric
Low
is less than
1. Re-plug the CO2 module.
2. If the problem persists, replace the
CO2 Hardware Error High
error
2. 12 V voltage error
3. Internal AD 2.5 V
error
4. Air pump error
1. Re-plug the CO2 module, or replace
the slot.
2. If the problem persists, replace the
CO2 module.
3. If the problem persists, replace the
module rack (refer to 11.1.14).
CO2 Sampleline
Occluded
Low
The sampling line is
occluded.
1. Check whether the sampling line is
blocked or folded.
2. If the problem persists, replace the
CO2 System Error Low
A multi-system error
occurs.
1. Re-plug the CO2 module, or replace
the slot.
2. If the problem persists, replace the
CO2 module.
3. If the problem persists, replace the
module rack (refer to 11.1.14).
CO2 No Watertrap
Low
The CO2 watertrap is off
or disconnected.
1. Check the CO2 watertrap.
2. Re-install the CO2 watertrap.
3. If the problem persists, replace the
CO2 watertrap.
4. If the problem persists, replace the
EtCO2 Overrange Low
exceeds the measurement
1. Recalibrate the CO2 module.
2. If the problem persists, replace the