7-43
FiCO2 Overrange Low
exceeds the measurement
1. Recalibrate the CO2 module.
2. If the problem persists, replace the
CO2 Zero Failed Low The CO2 module fails.
1. Re-zero the CO2 module.
2. If the problem persists, replace the
CO2 Init Error High
A CO2 initialization
error occurs.
1. Re-plug the CO2 module.
2. If the problem persists, replace the
CO2 module.
3. If the problem persists, check the
connection cable between the
backplane and the display.
4. If the problem persists, replace the
module rack (refer to 11.1.14).
5. If the problem persists, replace the
cable between the backplane and
Incompatible CO2
Software Version
High
incompatible with the
Replace the CO2 module (M02A
module not supported).
CO2 Sensor Error Low
The mainstream CO2
module sensor fails.
1. Re-plug the CO2 module.
2. If the problem persists, replace the
CO2 no sensor Low
module sensor is
disconnected, or a
communication error
1. Check the mainstream CO2 module
sensor.
2. Replace the mainstream CO2
module sensor.
CO2 Check Airway Low
airway is abnormal or
1. Re-plug the CO2 module.
2. If the problem persists, replace the
CO2 No Sampleline Low
No sampling line is
connected to the
MiniMedi CO2 module.
1. Check the sampling line of the CO2
module.
2. Replace the sampling line of the
CO2 module.
3. Replace the CO2 module.
CO2 Main Board
Error
High
The MiniMedi CO2
module fails.
1. Re-plug the CO2 module.
2. If the problem persists, replace the
CO2 Check Sensor
or Main Board
Low
The MiniMedi CO2
module fails.
1. Re-plug the CO2 module.
2. If the problem persists, replace the
CO2 Replace
Scrubber&Pump
Low
The MiniMedi CO2
module fails.
1. Re-plug the CO2 module.
2. If the problem persists, replace the
CO2 Replace Sensor Low
The MiniMedi CO2
module fails.
1. Re-plug the CO2 module.
2. If the problem persists, replace the
CO2 15V Voltage
Over Range
High
The MiniMedi CO2
module fails.
1. Re-plug the CO2 module.
2. If the problem persists, replace the
1. Re-plug the CO2 module.