7-29
Message
Priority
Cause
Solution
the external and internal AG
modules.
4. If both the measured values of the
external and internal AG modules in
step 3 are lower than 0.8% or higher
than 1.2%, replace the electronic
vaporizer.
5. If only the measured value of the
internal AG module in step 3 is
lower than 0.8% or higher than
1.2%, replace the internal AG
module (refer to 11.1.7).
7.3.6.1 AA Vaporizer Selftest Error Codes
Failure Code
Cause
FRU
Solution
00000001 Watchdog selftest error
Electronic
vaporizer
1. Reinstall the electronic
vaporizer.
2. Restart the machine.
3. If the problem persists, replace
the electronic vaporizer.
00000002 EEPROM selftest error
Electronic
vaporizer
1. Reinstall the electronic
vaporizer.
2. Restart the machine.
3. If the problem persists, replace
the electronic vaporizer.
00000004
Valve voltage error
Electronic
vaporizer
1. Reinstall the electronic
vaporizer.
2. Restart the machine.
3. If the power board has a voltage
error alarm, clear this alarm.
4. If the problem persists, replace
the electronic vaporizer.
00000008 Injector voltage error
00000010
Failed to open the safety valve
Electronic
vaporizer
1. Reinstall the electronic
vaporizer, or replace the slot.
2. Restart the machine.
3. If the problem persists, replace
the electronic vaporizer.
00000020 Failed to close the safety valve
00000040
Failed to open the drive gas
safety valve
Electronic
vaporizer
1. Reinstall the electronic
vaporizer, or replace the slot.
2. Restart the machine.
3. If the problem persists, replace
the electronic vaporizer.
00000080
Failed to close the drive gas
safety valve
7.3.6.2 AA Vaporizer Fault Error Codes
Failure
Code
Cause FRU Solution
00000001
Boil-off space
temperature too high
Electronic
vaporizer
1. Reinstall the electronic vaporizer, or
restart the machine.
2. If the problem persists, replace the
electronic vaporizer.

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA