7-30
Failure
Code
Cause FRU Solution
00000002
Boil-off space
temperature sensor
failure
Electronic
vaporizer
1. Reinstall the electronic vaporizer, or
restart the machine.
2. If the problem persists, replace the
electronic vaporizer.
00000004
Heating Module
Failure
Electronic
vaporizer
1. Reinstall the electronic vaporizer, or
restart the machine.
2. If the problem persists, replace the
electronic vaporizer.
00000008
Drive gas pressure
error in working
state
Electronic
vaporizer base
valve assembly
Secondary
pressure
regulator
Electronic
vaporizer
1. Replace the slot. If the alarm disappears,
replace the vaporizer base valve assembly
(refer to
11.1.12).
2. If the problem persists, insert another
vaporizer. If the fault alarm disappears,
replace the electronic vaporizer.
3. If the problem persists, replace the
pressure regulator of the electronic
vaporizer (refer to 11.1.4).
00000010
Drive gas pressure
error in standby state
Electronic
vaporizer base
valve assembly
Electronic
vaporizer
1. Replace the slot. If the alarm disappears,
replace the vaporizer base valve assembly
(refer to
11.1.12).
2. If the problem persists, insert another
vaporizer. If the fault alarm disappears,
replace the electronic vaporizer.
00000020
Failed to collect the
drive gas pressure
Electronic
vaporizer
Replace the electronic vaporizer.
00000040
Liquid level sensor
inconsistent with the
ultrasonic sensor
Electronic
vaporizer
Replace the electronic vaporizer.
00000080
Valve supply voltage
abnormal
Electronic
vaporizer
Replace the electronic vaporizer.
00000100
Injector supply
voltage abnormal
Electronic
vaporizer
Replace the electronic vaporizer.
00000200
Analog 5V voltage
abnormal
Electronic
vaporizer
Replace the electronic vaporizer.
00000400
Analog -5V voltage
abnormal
Electronic
vaporizer
Replace the electronic vaporizer.
00000800
CPU voltage
abnormal
Electronic
vaporizer
Replace the electronic vaporizer.
00001000
Injector limb failure:
The injection times
detected by the
injection detection
device are less than
the actual injection
times.
Electronic
vaporizer
Replace the electronic vaporizer.

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA