7-30
Cause FRU Solution
00000002
Boil-off space
temperature sensor
failure
Electronic
vaporizer
1. Reinstall the electronic vaporizer, or
restart the machine.
2. If the problem persists, replace the
00000004
Heating Module
Failure
Electronic
vaporizer
1. Reinstall the electronic vaporizer, or
restart the machine.
2. If the problem persists, replace the
00000008
Drive gas pressure
error in working
state
Electronic
vaporizer base
valve assembly
Secondary
pressure
regulator
Electronic
vaporizer
1. Replace the slot. If the alarm disappears,
replace the vaporizer base valve assembly
(refer to
11.1.12).
2. If the problem persists, insert another
vaporizer. If the fault alarm disappears,
replace the electronic vaporizer.
3. If the problem persists, replace the
pressure regulator of the electronic
vaporizer (refer to 11.1.4).
00000010
Drive gas pressure
error in standby state
Electronic
vaporizer base
valve assembly
Electronic
vaporizer
1. Replace the slot. If the alarm disappears,
replace the vaporizer base valve assembly
(refer to
11.1.12).
2. If the problem persists, insert another
vaporizer. If the fault alarm disappears,
replace the electronic vaporizer.
00000020
Failed to collect the
drive gas pressure
Replace the electronic vaporizer.
00000040
inconsistent with the
Electronic
vaporizer
Replace the electronic vaporizer.
00000080
Valve supply voltage
abnormal
Replace the electronic vaporizer.
00000100
Injector supply
voltage abnormal
Replace the electronic vaporizer.
00000200
Analog 5V voltage
abnormal
Replace the electronic vaporizer.
00000400
Analog -5V voltage
abnormal
Replace the electronic vaporizer.
00000800
Replace the electronic vaporizer.
00001000
The injection times
detected by the
injection detection
device are less than
the actual injection
Electronic
vaporizer
Replace the electronic vaporizer.