6-21
6.3.3.2 Zeroing Procedure
Perform the following steps to zero the EFCS.
1.Select Setup > System > Calibrate > Zero Flow Meters to access the screen shown below.
Select Begin to start zeroing.
2.The zeroing screen shown below is displayed after you select Begin. During zeroing, you can
select Cancel to abort the zeroing.