6-26
6. The screen shown below is displayed upon successful zeroing. Select Done to exit the
zeroing screen.
6.3.4.3 Zeroing Failure Troubleshooting
If the zeroing fails, perform the following steps to locate and rectify the fault.
1. Disconnect the gas supplies, exhaust the gas inside the machine, and perform zeroing again.
2. If the zeroing is successful, it is concluded that the previous zeroing failure is caused by a
mechanical failure in the pilot valve or faults in the proportional valves. Replace the pilot
valve assembly and proportional valve assemblies.
3. If the zeroing still fails, it is concluded that the cause is the fault in the Sensirion differential
pressure sensors or the high-flow O2 supply board. Replace the Sensirion differential
pressure sensor assemblies and high-flow O2 supply board.

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA