6-40
6.3.6.2 Calibration Procedure
Perform the following steps to calibrate the expiratory valve.
1. Enter the standby mode.
2. Select Setup > Service > Calibrate > Expiratory Valve to access the screen shown below.
3. Select Begin to start calibration. During calibration, you can select Cancel to abort the
calibration.