6-42
6. The screen shown below is displayed upon successful zeroing. Select Done to exit the calibration
screen.
6.3.6.3 Common Failures and Recommended Actions
Failure
Description
Possible Cause Recommended Action
Display E A zero point error occurs in the PEEP
sensor.
1. Calibrate the sensor.
2. Replace the VCM.
Display F The pressure change is nonlinear.
The larger the expiratory valve
sealing current, the smaller the
pressure.
1. Check for leakage.
2. Replace the drive gas assembly.
Display 10 The expiratory valve is nonlinear.
The larger the expiratory valve
sealing current, the smaller the
sampling current.
Replace the drive gas assembly.
Display 34
The drive gas pressure is low. 1. Replace or connect a gas supply to
keep the gas supply pressure within
280600 kPa (40 psi87 psi).
2. If the gas supply is normal, check the
gas supply pressure switch.

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA

Related product manuals