6-57
3. Select Setup > Service > Calibrate > Vaporizer Injector to access the screen shown below.
Select vaporizer 2 (for example, Iso vaporizer). (Note: The calibration steps of vaporizer 1 are the
same as those of vaporizer 2.)
4. Select Begin to start calibration. During calibration, you can select Cancel to abort the
calibration.