7-35
Message
Priority
Cause
Solution
halogenated agents. The
measured MAC is less
than 3.
Mixed Agent Med
The AG module detects
two and more types of
halogenated agents. It
cannot detect the
specific MAC value
(invalid value).
1. Check whether only one type of
halogenated agent is used.
2. If only one type of anesthetic agent is
used, set the fresh gas flow to 6L/min or
above, and check 20s later.
3. If the problem persists, replace the AG
module.
Mixed Agent and
MAC ≥ 3
Med
The AG module detects
two and more types of
halogenated agents. The
measured MAC is
greater than or equal to
3.
1. Check whether only one type of
halogenated agent is used.
2. If only one type of anesthetic agent is
used, set the fresh gas flow to 6L/min or
above, and check 20s later.
3. If the problem persists, replace the AG
module.
CO2 Overrange Low
The monitored value
exceeds the
measurement range of
the module.
1. Decrease the concentration of the
monitored gas to the normal range.
2. If the problem persists, calibrate the
AG module again.
3. If the problem persists, replace the AG
module.
O2 Overrange Low
The monitored value
exceeds the
measurement range of
the module.
1. Decrease the concentration of the
monitored gas to the normal range.
2. If the problem persists, calibrate the
AG module again.
3. If the problem persists, replace the AG
module.
N2O Overrange Low
The monitored value
exceeds the
measurement range of
the module.
1. Decrease the concentration of the
monitored gas to the normal range.
2. If the problem persists, calibrate the
AG module again.
3. If the problem persists, replace the AG
module.
N2O Overrange Low
The monitored value
exceeds the
measurement range of
the module.
1. Decrease the concentration of the
monitored gas to the normal range.
2. If the problem persists, calibrate the
AG module again.
3. If the problem persists, replace the AG
module.
Hal Overrange Low
The monitored value
exceeds the
measurement range of
the module.
1. Decrease the concentration of the
monitored gas to the normal range.
2. If the problem persists, calibrate the
AG module again.
3. Replace the AG module.
Enf Overrange Low
The monitored value
exceeds the
measurement range of
the module.
1. Decrease the concentration of the
monitored gas to the normal range.
2. If the problem persists, calibrate the
AG module again.
3. Replace the AG module.
Iso Overrange
Low
The monitored value
1. Decrease the concentration of the

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA

Related product manuals