7-39
Message
Priority
Cause
Solution
BIS sensor.
BIS Electrode 3 Lead
Off
Low
BIS electrode 3 falls off
from the patient.
1. Check the contact between BIS
sensor electrode 3 and the patient skin.
2. If the problem persists, replace the
BIS sensor.
BIS Electrode 4 High
Imped.
Low
The impedance of BIS
electrode 4 is too high.
1. Check the contact between BIS
sensor electrode 4 and the patient skin.
2. If the problem persists, replace the
BIS sensor.
BIS Electrode 4 Lead
Off
Low
BIS electrode 4 falls off
from the patient.
1. Check the contact between BIS
sensor electrode 4 and the patient skin.
2. Replace the BIS sensor.
BIS Electrode G High
Imped.
Low
The impedance of BIS
electrode G is too high.
1. Check the contact between BIS
sensor electrode G and the patient
skin.
2. If the problem persists, replace the
BIS sensor.
BIS Electrode G Lead
Off
Low
BIS electrode G falls off
from the patient.
1. Check the contact between BIS
sensor electrode G and the patient
skin.
2. If the problem persists, replace the
BIS sensor.
BIS Electrode C High
Imped.
Low
The impedance of BIS
electrode C is too high.
1. Check the contact between BIS
sensor electrode C and the patient
skin.
2. If the problem persists, replace the
BIS sensor.
BIS Electrode C Lead
Off
Low
BIS electrode C falls off
from the patient.
1. Check the contact between BIS
sensor electrode C and the patient
skin.
2. Replace the BIS sensor.
BIS Electrode LE
High Imped.
Low
The impedance of BIS
electrode LE is too high.
1. Check the contact between BIS
sensor electrode LE and the patient
skin.
2. If the problem persists, replace the
BIS sensor.
BIS Electrode LE
Lead Off
Low
BIS electrode LE falls
off from the patient.
1. Check the contact between BIS
sensor electrode LE and the patient
skin.
2. If the problem persists, replace the
BIS sensor.
BIS Electrode LT
High Imped.
Low
The impedance of BIS
electrode LT is too high.
1. Check the contact between BIS
sensor electrode LT and the patient
skin.
2. If the problem persists, replace the
BIS sensor.
BIS Electrode LT
Lead Off
Low
BIS electrode LT falls
off from the patient.
1. Check the contact between BIS
sensor electrode LT and the patient
skin.
2. If the problem persists, replace the
BIS sensor.
BIS Electrode RE
Low
The impedance of BIS
1. Check the contact between BIS

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA

Related product manuals