2-36
EFCS flowmeter monitoring board (bottom view)
2.2.5.2 Flow Sensor Board
The flow sensor board is mainly used to monitor the O2/N2O/air flow and feed back the detected
signals to the EFCS flowmeter monitoring board through the I2C bus.
Installation position of the flow sensor board
Flow sensor board (top view) Flow sensor board (side view)
2.2.5.3 BFCS Switch Board
The BFCS switch board is mainly used to monitor the open/closed status of the BFCS door. Two
Hall switches are used to feed back the status signals of the BFCS door to the EFCS flowmeter
monitoring board.
Flow sensor board

Table of Contents

Other manuals for Mindray A9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray A9 and is the answer not in the manual?

Mindray A9 Specifications

General IconGeneral
TypeAnesthesia Workstation
Integrated Gas AnalysisYes
Backup BatteryYes
Operating Temperature10°C to 40°C
Storage Temperature-20°C to 55°C
Battery TypeLithium-ion
Ventilation ModesVolume Control, Pressure Control, SIMV, Manual Ventilation
DisplayTouchscreen
Patient MonitoringECG, SpO2, NIBP, EtCO2, Temperature
ConnectivityEthernet, USB
PortabilityMobile (with wheels)
Battery LifeMinimum 90 minutes (typical)
AltitudeUp to 3000 meters
Power Supply100-240 VAC, 50/60 Hz
Charging TimeApproximately 4 hours
StorageInternal storage for patient data and system logs
Supported LanguagesMultiple languages
CertificationsCE, FDA