7 Service and Maitenance 7-9
5
CAUTION
The tweezers may scratch the probe. Exercise caution
when using it to clean the probe. Avoid direct contact
between the tweezers and the probe. Do not use
excessive force when cleaning the probe. Otherwise it
may bend.
NOTE
Mindray recommends the acid and alkaline detergents be
used alternately for this purpose. For instance, if the acid
detergent has been used for last maintenance, the
alkaline detergent had better be used for this time.
Use tweezers to pinch acid or alkaline detergent-soaked gauze, and
then gently clean the exterior of the probe until it is clean and smooth.
6
Use tweezers to pinch deionized water-soaked gauze to clean the
probe.
7
After cleaning, gently pull the probe arm to its highest point and rotate
the probe arm to move the probe to a position above the wash pool.
8
Load the sample/reagent disk.
9
Install the sample/reagent disk cover.
7.3.2 Cleaning Mixing Bar
WARNING
The probe tip is sharp and can cause puncture wounds. To
prevent injury, exercise caution when working around the probe.
BIOHAZARD
In case your skin contacts the sample, control or calibrator, follow
laboratory safety procedure and consult a doctor.
Dispose of the used gauze in accordance with your local or national
guidelines for biohazard waste disposal.
1
Place the Power to OFF.
2
Gently pull the bar arm to its highest point and rotate it to move the bar to
a position convenient to operate.