7 Service and Maitenance 7-
15
7.4.3 Cleaning Sample/Reagent Rotor
WARNING
The probe tip is sharp and can cause puncture wounds. To prevent
injury, exercise caution when working around the probe.
BIOHAZARD
Wear gloves and lab coat and, if necessary, goggles.
Dispose of the used gauze in accordance with your local or national
guidelines for biohazard waste disposal.
1 Place the Power to OFF.
2 Pull the probe arm to its highest point.
3 Wipe the rotor with clean gauze
7.5 Three-month Maintenance
7.5.1 Washing Dust Screen
1
Place the Main Power to OFF.
2
Use a screwdriver to unscrew the screws on the right plate and remove
the right plate.You can see the front and back dust screen on the right
side.The dust screen is fixiated by the bullet and stop plate.