9 Troubleshooting
9-12
400001360000
Test result error: Blank
response of … too high
Blank reponse too high Check the cuvette, reagent and diluent.
400001370000
Test result error:
Calibration sensitivity
of … too low
Reagent blank ABS too high
or decreased ABS resulting
from calibrator degradation
Refill the reagent and replace calibrator; recalibrate.
500000170000
System environment
error: Printer connection
error
Printer connection error
Check the connection between the printer and the
operation unit and make sure the printer is turned on.
500001160000
Operating software
error: is out of date.
Please replace in time
Expired Replace the software.
500001180000
Operating software
error: Calibration interval
of is out. Please
re-calibrate in time
Calibration interval reached
Recalibrate.
9.4 Analyzing Unit Error
9.4.1 Main Unit
Error Code Error Message Reason Corrective Measure
100640010007
Main unit result error:
Command error
Main unit receiving
command frame error
Shut down Power and restart the PC and open the
operating software. Execute failure recovery. Please contact
the R&D engineer if the error repeats for 3 times.
100640020007
Main unit result error:
Self-check error
Self-check error
Shut down Power and restart the PC and open the
operating software. Execute failure recovery. Please contact
the R&D engineer if the error repeats for 3 times.
100640030007
Main unit result error:
Shaking hands with
other units
Shaking hands with other
units
Execute the previous operation after waiting for 30-60
seconds. If there is no response for a long time, execute
mechanical resetting and check whether the mainboard
connection is ok.