2 System Performance and Workflow
2-4
Reagent volume
30µl-450µl, with increment of 1µl
Reagent position assignment
Sample and reagent share one sample/reagent disk which is of single-circle
structure. No.9-No.36 are reagent positions and No. 35 is fixed for wash solution
and No. 36 is fixed for dilution.Other positions can be assigned for R1 or R2. If
ISE module is installed, No.34 is fixed for ISE wash solution.
Reagent input mode
Enter manually
When hand-held bar code system is installed, the reagent information can be
entered using the bar code. Refer to the documents accompanying the optional
hand-held bar code reader.
Reagent probe
Sample and reagent share one probe, which is capable of detecting liquid level,
protecting the probe against collision in the vertical direction and tracking liquid
level.
Reagent probe washing
Inside and outside of the probe are washed with carryover less than 0.1%.
2.1.4 Specifications of Reaction System
Optical path of reaction cuvette
5mm
Material of reaction cuvette
5mm×6mm×30mm, disposable reaction cuvette
Number of reaction cuvettes
40
Mixing method
One mixing bar, which starts to stir after adding a samples or R2
Photometric System
Filter wheel optical system
Wavelength
8 wavelengths:340nm、405nm、450nm、510nm、546nm、578nm、630nm、
670nm
Wavelength accuracy
±2nm