3 Installation  3-
Figure 3-4 Remove the Fixing Plate 
 
3.3.2    Installation 
  Fix  the analyzer: after the analyzer is placed  on the target installation site, 
adjust the  two front fixing bolts to  ensure the  four of  them have  the same 
height (the two behind are not adjustable).   
  Remove the plastic cover and fixing bandage, and install the sample probe 
and  mixing  bar.  Do  not  install  the  arm  cover  of  the  sample  probe  before 
calibrating the level detection board.   
•  Place the ANALYZING UNIT POWER to ON while ensuring that the sample 
probe is not attaching any conducting .object, such as hands.   
•  Calibrate  the  sample  probe  manually.  Check  if  indicator  D2  (yellow)  is 
lightened within 2 seconds when the ANALYZING UNIT POWER to ON. Press 
the switch S2 on the level detection board and then release it, then check if 
indicator D2 is first extinguished and then lightened. If it is, that means the 
calibration succeeds.   
Exercise caution to prevent the sample probe from attaching any conducting 
object during the calibration. 
•  Place the ANALYZING UNIT POWER to OFF.   
NOTE:
::
: 
 
Remember to install the washer when installing the sample probe. 
 
Use syringe to inject water into the filter and connect the filter to the cap assembly of 
the DI water tank. Connect the liquid tubes as indicated in the following figure. 
 
Remove  the 
package 
plates  of  the 
front and back 
feet.