3 Installation 3-3
The distance between the power socket and the system should be less than
3 meters.
WARNING
Make sure the power socket is grounded correctly. Improper
grounding may lead to electric shock and/or equipment damage.
Be sure to connect the system to a power socket that meets the
requirements mentioned above and has a proper fuse installed.
3.2.3 Water Supply and Drainage Requirements
The supplied water must meet requirements of the CAP Type II water, with
specific resistance no less than 0.5(M.cm@25).
The water temperature should be within 5-32.
BIOHAZARD
Dispose of waste liquids according to your local regulations.
CAUTION
The supplied water must meet requirements of the CAP Type II
water; otherwise insufficiently-purified water may result in
misleading measurement.
3.3 Installation Procedures
3.3.1 Unpacking
Check the delivery list before unpacking. Besides PC and printer box, there
are three boxes for analyzing unit, accessory and deionized water tank,
waste tank and used-cuvette bucket.
The gross weight of the analyzing unit is about 95Kg. 3-4 people are needed
to lay the wooden case containing the analyzing unit on the gound. Use fork
truck, if possible.
Use special tools to unpack the top cover and the side plate.