8 Test and Maintenance Software
8-7
Figure 8-6 Sample Unit Debugging Interface
8.2.5 Reaction Unit
Click Reaction Unit on the higher-left area and then you can see the fuction buttons of
the reaction unit.
Reaction tray rotate——Input the number of the circles to rotate in the first edit box
and the position to stop in the second edit box, which means to stop the specified
cuvette to the 1# dispensing position of the reaction disk.
When you input “29” into the second edit box, the first cuvette segment will stop at the
replacement position; input “34”, the second cuvette segment will stop at the replacement
position; input “39”, the third cuvette segment will stop at the replacement position; input
“4”, the fourth cuvette segment will stop at the replacement position…and the rest cuvette
segment can be deduced by analogy.
Reaction tray rotate some cupsThe number of cuvettes the reaction disk
passes while rotating.
Turn on light
Turn off light
Rotate and measure—— Collect photoelectric data of all the cuvettes on the reation
disk at all wavelengths.