3 Installation 3-2
Do not use devices such as mobile phones or radio transmitters near the
system. Electromagnetic waves generated by those devices may interfere
with operation of the system.
The altitude height of the installation site should be lower than 2000 meters.
Ambient temperature: 15℃-30℃, with fluctuation less than ±2℃/H.
Relative humidity: 35%RH-85%RH, without condensation
CAUTION
Operating the system in an environment other than the specified
may lead to unreliable test results.
If the temperature or relative humidity does not meet the
requirements mentioned above, be sure to use air-conditioning
equipment.
3.2 Installation Requirements
3.2.1 Space and Accessibility Requirements
Figure 3-1 Space and Accessibility Requirements
3.2.2 Power Requirements
Power supply: 100-130V/200-240V, 50/60Hz, three-wire power cord and
properly grounded.
The system should be connected to a properly grounded power socket.