2 System Performance and Workflow
2-2
Reaction liquid volume
BS-120/BS-130: 180-500µl
BS-180/BS-190: 150-500µl
Reaction temperature
37℃
Test scope
Clinical chemistries, immunoassays, TDM (Treatment Drug Monitoring)
Auto dilution
Dilution ratio 4~150; dilution is done in reaction cuvette.
Operation mode
System and tests are configured via the operating software. Profiles and
calculation tests are allowed.
Calibration method
Linear (one-point, two-point and multi-point), Logit-Log 4p, Logit-Log 5p, Spline,
Exponential, Polynomial and Parabola
Programming Controls
Westgard Multi-rule, Cumulative sum check, Twin plot
Data processing
Capable of storing and outputting various data and tables/graphs, and calculating
among different tests
Dimensions
l×b×h:690 mm×570 mm×595 mm.
Weight
75kg
Emergent samples
Emergent samples can be inserted during test at any time.
Network connection
Able to be connected with LIS (Laboratory Information Management System)
2.1.2 Specifications for Sample System
Sample tube type
Microtube: Φ10×37mm, Φ12×37mm;
Blood collecting tube: Φ12×68.5mm, Φ12×99, Φ12.7×75, Φ12.7×100, Φ13×75,
Φ13×100;