EasyManuals Logo

Mindray BS-130 User Manual

Mindray BS-130
267 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #28 background imageLoading...
Page #28 background image
2 System Performance and Workflow
2-3
Plastic tube: Φ12×68.5mm, Φ12×99, Φ12.7×75, Φ12.7×100, Φ13×75,
Φ13×100.
Minimum sample volume
Minimum sample volume=dead volume of the sample+total sample volume
needed for all the tests
Dead volume of the sample: Microtube 300, Blood collecting tube500ul, Plastic
tube500ul.
Sample position
Sample and reagent share one sample/reagent disk which is of single-circle
structure. No.1-No.8 are sample positions which can be set as routine, QC, STAT
positions.
STAT sample
Emergent samples can be inserted during test at any time and then run with high
priority.
Sample volume
3µl-45µl, with increment of 0.5µl
Sample probe
Sample and reagent share one probe, which is capable of detecting liquid level,
protecting the probe against collision in the vertical direction and tracking liquid
level.
Sample probe washing
Inside and outside of the probe are washed with carryover less than 0.1%.
Sample input mode
Enter manually
When hand-held bar code system is installed, the sample information can be
entered by using the bar code. Refer to the documents accompanying the
optional hand-held bar code reader.
2.1.3 Specifications for Reagent System
Reagent refrigeration
24 hours non-stop refrigeration, refrigeration temperature: 4-15
Reagent dispensing
Reagent is aspirated and dispensed precisely by syringes.
Reagent types
1 to 2 reagent types, R1 and R2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray BS-130 and is the answer not in the manual?

Mindray BS-130 Specifications

General IconGeneral
BrandMindray
ModelBS-130
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals