7-14 7 Service and Maitenance
WARNING
The probe tip is sharp and can cause puncture wounds. To prevent
injury, exercise caution when working around the probe.
BIOHAZARD
Wear gloves and lab coat and, if necessary, goggles.
Dispose of used cotton swabs in accordance with your local or
national guidelines for biohazard waste disposal.
1
Place the Power to OFF.
2
Pull the sample probe arm to the highest point by hand. Rotate the bar
arm to move the probe to a position convenient to operate.
3
Clean the inside of and the place around the wash pool with cotton
swabs.
4
Rotate the sample probe back to a position above the wash pool.
7.4.2 Cleaning Wash pool of Mixing Bar
WARNING
The probe tip is sharp and can cause puncture wounds. To prevent
injury, exercise caution when working around the probe.
BIOHAZARD
Wear gloves and lab coat and, if necessary, goggles.
Dispose of the used cotton swabs in accordance with your local or
national guidelines for biohazard waste disposal.
1 Place the Power to OFF.
2 Gently pull the bar arm to its highest point and rotate it to move the bar to
a position convenient to operate.
3 Clean the inside of and the place around the wash pool with cotton
swabs.
4 Pull the mixing bar arm to its highest point and rotate it to move the bar to
a position above the wash pool.