8-22 8 Test and Maintenance Software
Figure 8-19 Dark Current Test Interface
8.5.5 Other Functions of the Photoelectric Test
In the Photoelectric tab, you can also test the basic performance, including absorbance
repeatability, precision, stability, drift, stray light and carryover. Since they are not included
in the maintenance procedure, we will not introduce them in this manual.
8.6 Macro Instructions
8.6.1 Function
Each unit tab of the test and maintenance software only includes the basic commands
and procedures for ordinary alignment, which do not support more complex and
comprehensive tests. However, the Macro Instruction tab allows you to create necessary
instructions or instruction set (macroinstruction) for the tests.
8.6.2 Detailed Operations
8.6.2.1 Screen Layout
Click the Setup button on the higher-right area shown in Figure 8-13, input “bs120” to the
dialog box and then click OK. You will enter the administrator interface shown in Figure
8-20.
On the right side of the screen is the Macro Instruction area, in which you can create and
edit macro instructions and execute them circularly.
On the left side of the screen is the instruction area, in which you can view the detailed
instructions of the specified unit.