7-30 7 Service and Maitenance
2
Install the probe by following the procedure in 7.6.4.3 Installing Probe.
CAUTION
Pull the probe to the above of wash pool after installing it
and then load the sample/reagent disk.
7.6.6 Replacing Mixing Bar
If the mixing bar is damaged, it must be replaced immediately. Follow the procedure
given below to replace the damaged mixing bar.
WARNING
The probe tip is sharp and can cause puncture wounds. To prevent
injury, exercise caution when working around the probe.
When replacing the bar, pinch the bar only by the knurled part and do
not touch the other part of the bar. Protect the flat part of the bar from
being scratched.
BIOHAZARD
Wear gloves and lab coat and, if necessary, goggles.
Dispose of the damaged mixing bar in accordance with your local or
national guidelines for biohazard waste disposal.
CAUTION
Please use Mindray-recommended consumables. Other
consumables may decrease the system performance.
1
Place the Power to OFF.
2
Gently pull the bar arm to a position convenient to operate.