9 Troubleshooting 9-25
9.4.5 Mixing Unit
Error Code Error Message Reason Corrective Measure
100670010005
Mixing unit result error:
Command error
Command sending error
or wrong command
information
Reset the Power and restart the operating software, then
retry this command. If the message repeats continuously for
3 times, please contact the R&D engineer.
100670020005
Mixing unit result error:
Self-check error
Self-check error
Reset the Power and restart the operating software and
retry this command. If the message repeats continuously for
3 times, please contact the R&D engineer.
100670030005
Mixing unit result error:
Mechanical resetting
error
Mechanical resetting error
Reset the mechanical parts by using Alignment. If the
warning remains, check the connection of the motor and the
sensor.
100670040005
Mixing unit result error:
Status error. Self-check
Self-checking; Other
operations not supported
Execute the previous operation after waiting for 30-60
seconds. If there is no response for a long time, execute
mechanical resetting. If the error repeats for several times,
please contact the R&D engineer.
100670040015
Mixing unit result error:
Status error. Error
Mixing unit error; current
operations not supported
Execute failure recovery. If the warning remains, check the
connection of the motor and the sensor.
100670040025
Mixing unit result error:
Status error. Waiting for
handshake
Unit busy. No response
Execute the previous operation after waiting for 30-60
seconds. If there is no response for a long time, execute
mechanical resetting. If it repeats for several times, please
contact the R&D engineer.
100670050005
Mixing unit result error:
Unit busy. No response
Unit busy. No response
Execute the previous operation after waiting for 30-60
seconds. If there is no response for a long time, execute
mechanical resetting. If it repeats for several times, please
contact the R&D engineer.
100670080005
Mixing unit result error:
Undefined configuration
Undefined configuration Configure the right parameter.
100670090005
Mixing unit result error:
Wrong configuration
parameter
Wrong configuration
parameter
Configure the right parameter.