9 Troubleshooting
9-24
100660080000
Temperature unit result
error: Undefined
temperature parameter
Wrong system operation
parameter
Configure the right parameter.
100660090000
Temperature unit result
error: Wrong
temperature parameter
Wrong system operation
parameter
Configure the right parameter.
100660100000
Temperature unit result
error: Undefined sensor
parameter
Undefined sensor
parameter
Configure the right parameter.
100660110000
Temperature unit result
error: Wrong sensor
parameter
Wrong sensor parameter Configure the right parameter.
100660120000
Temperature unit result
error: Undefined target
temperature
Undefined target
temperature
Configure the right parameter.
100660130000
Temperature unit result
error: Wrong target
temperature parameter
Wrong target temperature
parameter
Configure the right parameter.
100660160000
Temperature unit result
error: Parameter write
protection
Unit parameter write
protection. Parameter
configuration forbidden
Configure the parameter after parameter configuration is
permitted.
100661250000
Temperature unit result
error: Main unit sending
error
/
Check the connection of the temperature unit. Restart the
operating software and retry the command. If the error
repeats for several times, check the cable and the board.
100661270000
Temperature unit result
error: Main unit does
not receive the
Temperature unit result
/
Check the connection of the temperature unit. Restart the
operating software and retry the command. If the error
repeats for several times, check the cable and the board.