9 Troubleshooting
9-14
100640170007
Main unit result error:
Analyzing error.
Reaction disk affected
Other units status error.
Photoelectric collection
affected
Restart the analyzing unit and open the operating software.
Execute failure recovery. If the error repeats continuously
for 3 times, please contact the R&D engineer.
100640180006
Main unit result error:
Analyzing error.
Reaction disk not
affected
Other units status error.
Photoelectric collection
not affected
After the tests are finished, restart the analyzing unit and
open the operating software. Execute failure recovery.If the
message appears continuously for 3 times, please contact
the R&D engineer.
100640300007
Main unit result error:
Parameter write
protection
Parameter write protection
Parameter write protection is permitted.
100640310007
Main unit result error: No
result, or timeout
System connection error
or other units error
Check the connection of the system; restart the analyzing
unit and open the operating software. Execute failure
recovery.
9.4.2 Sample/Reagent Unit
Error Code Error Message Reason Corrective Measure
100680010005
Sample unit result error:
Command error
Commamd sending error
or wrong command
information
Restart the analyzing unit and rerun the operating software.
If the error repeats continuously for 3 times, please contact
the R&D engineer.
100680020005
Sample unit result error:
Self-check error
Self-check error
Restart the analyzing unit and rerun the operating software.
If the error repeats continuously for 3 times, please contact
the R&D engineer.
100680030005
Sample unit result error:
Mechanical resetting
error
Mechanical resetting error
Reset the mechanical parts by using Aligment. If the error
remains, check the connection of the motor and the sensor.
100680040005
Sample unit result error:
Status error. Self-check
Self-checking; other
operations not supported
Execute the previous operation after waiting for 30-60
seconds. If there is no response for a long time, execute
mechanical resetting. If the error remains, restart the
analyzing unit and the operating software.