9 Troubleshooting   
9-2 
 
9.2 Classification of Error Messages 
Severity: Warning 
Level  Description  Actions taken by the system 
0  Errors to neglect  The system only reminds you of the errors and will not take any actions. 
1  Errors to flag tests  The system flags the tests
 
when abnormity of the system occurs or the unreliable results are 
generated due to any reasons during the test 
2  Errors  to  invalidate 
sample 
When  a  test  is  invalidated  due  to  the  abnormal  sample,  the  system  will  rerun  the  test 
immediately. 
3  Errors to skip sample  Samples for certain tests are used out. 
The system ignores all tests related to the sample and continues with other tests. You can 
refill the sample and resume the tests after other tests are finished, or after clicking the Probe 
Stop button. 
4  Errors  to  invalidate 
reagent 
When  a  test  is  invalidated  due  to  the  erroneous  reagent,  the  system  will  rerun  the  test 
immediately. 
5  Errors to skip reagent
Reagents for certain tests are used out. 
The system skips all tests related to the reagent and continues with other tests. You can refill 
the reagent and resume the tests after other tests are finished, or after clicking the Probe 
Stop button. 
6  Errors  to  invalidate 
sample/reagent 
The system will invalidate all tests that are related to the reagent and sample. 
 
Severity: Pausing 
Level  Description  Actions taken by the system 
7  Errors  to  pause  The system will pause the probe/mixing bar and invalidate all tests but those which already